HORIBA
ASGU-370TS SPLIT
Automatic Span Gas Generator Unit for Quality control at field test
Specifications:
| Principle | Dilution (incl. built in cylinder), Permeation, Gas Phase Titration, (Mixed gas dilution)  | 
| Application | Quality control in LaboratoryQuality control in field
 | 
| Stage of extension | Only one channel (module) per unit | 
| Flow regulation | Digital Mass Flow Controllers | 
| Span gas repeatability | +/- 1% | 
| Span gas linearity | +/- 2% (minimum flow > 20% full scale) | 
| Permeation System (NO2, SO2, NH3, H2S, BTX) | PID temperature controller with Pt 100 sensorTemperature: 50 °C
 Accuracy ± 0,1 °C
 Indication of actual temperature on display
 | 
| Zero Air Supply | Internal Zero gas preparation | 
| Operation | Manual (Touch Screen Display)External via RS232 (Bayern Hessen protocol)
 External vie Ethernet (Remote Software or Bayern Hessen protocol)
 | 
| Battery pack | For uninterruptible power supply for approx. 30 min. purge mode without mains voltage | 
| Power | 230 V / 50HzPower consumption depends on channel
 | 
| Case 1(Span Gas Generator)
 | Span Gas Source (Span Gas Cylinder or Permeation Tube)Span Gas Generation (Bronkhorst TMDs)
 Controlling Software
 Datalogger (Option)
 UPS – Battery Modul (Option)
 Dimension: 600mm(W) x 400mm(H) x 155mm(D)
 Weight depends on stage of extension: 12 - 16kg
 |