Specifications:
| General Spectrometer Specifications | 
| Spectral Coverage | 115 - 230 nm Full coverage on CCD: 115 - 320 nm (No order sorting filter available) | 
| Spectral Resolution  | 0.5 nm with 22 μm pinhole, 0.75 nm with 40 μm pinhole | 
| Spectral Dipersion | 12 nm/mm; 0.17 nm/pixel | 
| Focal Length  | 90 mm | 
| VUV Optics | MgF2 coated grating and mirror | 
| Vacuum or Nitrogen Purged Mode | ~ 10-6 Torr with turbomolecular pump. Helium leak test performed at factory.N2 high purity nitrogen constant purge required
 | 
| Options
 | Selection of different heights of VUV CCDsKinematic mounts to interchange Pinholes (or Slits)
 Bare metal finish is standard: Super black coating reduces stray light
 | 
| F/#  | ~ F/10 | 
| Software | LabVIEW Acquisition sortware for initial evaluation (DLLs provided for software integration) | 