| Number of channels | 1 channel/unit | 
| Bridge resistance | 60 to 1,000 Ω | 
| Gauge factor | 2.00 | 
| Bridge power supply (BV) | Voltage | 0.5 V, 2 V AC | 
| Frequency | 5 kHz sine wave | 12.5 kHz sine wave | 25 kHz sine wave | 
| Synchronization input/output signal (OSC) | 2.5 V AC | 
| E-terminal Connection Switching function | The Shield Wire Potential Switching switch (E-SW) can be used to switch the potential of the E-terminal (Shield Wire) of the input cable to the input common (COM) or GND (casing ground potential via protective element). | 
| Bridge Check (BRC) function | Detects disconnections and shorts in input bridge circuits with bridge resistance of 120 Ω or more and disconnections in input cables, and displays the results on LEDs. Function can be turned ON/OFF by FNC setting. | 
| Cable Length Compensation (CLC) function | Automatic compensation for bridge power supply voltage drop due to cable length to bridge circuit (bridge resistance 120 Ω or more). Function can be turned ON/OFF by FNC setting. | 
| Balancing adjustment range(Auto-balance) | Resistance Deviation: ±2% (±10,000 x 10-6 strain), capacitance imbalance: Approx. 2,000 pF | 
| Balancing adjustment accuracy | Within ±0.4 ×10-6 strain(RANGE = 200, FINE = x1, BV = 2 V)
 | Within ±1.0 ×10-6 strain(RANGE = 500, FINE = x1, BV = 2 V)
 | Within ±1.0 ×10-6 strain(RANGE = 500, FINE = x1, BV = 2 V)
 | 
| Maximum input range | ±200,000 x 10-6 strain(RANGE = 20k, FINE = x2.5, BV = 0.5 V)
 | ±500,000 x 10-6 strain(RANGE = 50k, FINE = x2.5, BV = 0.5 V)
 | ±500,000 x 10-6 strain(RANGE = 50k, FINE = x2.5, BV = 0.5 V)
 | 
| Voltage sensitivity | ±10 V output at ±200 ×10-6 strain (RANGE = 200, FINE = x1, BV = 2 V) | ±10 V output at ±500 ×10-6 strain (RANGE = 500, FINE = x1, BV = 2 V) | ±10 V output at ±500 ×10-6 strain (RANGE = 500, FINE = x1, BV = 2 V) | 
| Measurement range (RANGE) | 200, 500, 1k, 2k, 5k, 10k, 20k (x10-6strain × 2/BV value), OFF | 500, 1k, 2k, 5k, 10k, 20k, 50k (x10-6strain × 2/BV value), OFF | 500, 1k, 2k, 5k, 10k, 20k, 50k (x10-6strain × 2/BV value), OFF | 
| Measuring range fine adjustment (FINE) | Continuously variable within the RANGE and adjustable with the FINE knob. Variation can be changed in two steps. | 
| Internal Calibrator (+CAL, -CAL) | Calibration value range: ±1 to 9,999 x10-6 strain, calibration value accuracy: ±(0.5% rdg + 0.5 ×10-6 strain) | 
| Auto-range (ARG) function | Automatic adjustment of RANGE and FINE so that the output voltage becomes the specified value when the calibration value set in the Internal Calibrator is applied.Output voltage value (OFF, 5 V, 8 V, 10 V) can be specified by the FNC setting.
 | 
| Non-linearity | ±0.1%/FS | ±0.1%/FS | ±0.2%/FS | 
| Frequency characteristic (W/B) | DC to 2 kHz ±10% | DC to 5 kHz ±10% | DC to 10 kHz ±10% | 
| High-Pass Filter (HPF) | Cutoff frequency: 0.5 Hz, 2-pole Butterworth type (filter drop characteristic -12dB/oct) | 
| Low-Pass Filter (LPF) | Cutoff frequency: 10, 30, 100, 300, 500 Hz, 4-pole Butterworth type (filter drop characteristics -24dB/oct) | Cutoff frequency: 10, 30, 100, 500, 3 kHz, 4-pole Butterworth type (filter drop characteristics -24dB/oct) | Cutoff frequency: 10, 30, 100, 500, 3 kHz, 4-pole Butterworth type (filter drop characteristics -24dB/oct) | 
| Temperature stability | Sensitivity | Within ±0.02%/°C, Within ±0.2%/24h | 
| Zero-point | Within ±0.1 ×10-6 strain/°C, ±0.5 ×10-6 strain/24h | 
| Noise | 2.0 ×10-6 strain p-p input conversion(W/B, RANGE = 200,
 FINE = x1, BV = 2 V, at 120 Ω bridge)
 | 5.0 ×10-6 strain p-p input conversion(W/B, RANGE = 500,
 FINE = x1, BV = 2 V, at 120 Ω bridge)
 | 7.0 ×10-6 strain p-p input conversion(W/B, RANGE = 500,
 FINE = x1, BV = 2 V, at 120 Ω bridge)
 | 
| Output voltage | OUTPUT1 ±10 V ±5 mA (2 kΩ load or more), OUTPUT2 ±10 V ±10 mA (332 kΩ load or more), Operable at output resistance 0.5 Ω or less, capacitive load up to 0.1 µF. | 
| Output adjustment | OUTPUT2 ADJ (continuously variable from 1 to 1/10 independently) | 
| Level indication | 17-dot LED display for monitoring OUTPUT1, Green LED in the center lights up within approx. ±100 mV and LEDs on both ends blink at approx. ±10.5V or more | 
| Digital display | 4 1/2 digit digital display for OUTPUT2 monitor, scaling available with OUTPUT2 ADJ. Decimal point position can be changed by FNC setting.Accuracy: Within ±0.05% rdg ±1 count | 
| Key Lock function | Key lock ON/OFF of the operation panel is possible by pressing the key lock button for about 1 second (excluding the calibration value application switch (+CAL, -CAL) ) | 
| Remote function | Rear interface connector: D-SUB, 9P, female, Mounting: #4-40 inch screw+CAL and -CAL input, BAL function [Bridge check (BRC), Line length correction (CLC), Auto range (ARG), Auto balance (resistance balance)], RANGE OFF and KEY LOCK controls are possible by external signal
 | 
| Saving of setup values | Save to flash memory | 
| Vibration resistance | 29.4m/s2 (50 Hz, XYZ each for 10 minutes) and compliant with MIL-STD-810G-514.6 | 
| Withstand voltage | Between input terminals (A, B, C, D, E) and the output, and the case: 1 kV AC for 1 minuteBetween AC power input and signal input, output, and the case: 1.5 kV AC for 1 minute
 Between DC power input and signal input: 1 kV AC for 1 minute
 Between DC power input and signal output, and the case: 500 V AC for 1 minute
 | 
| AC power supply | 100 to 120 V AC / 200 to 240 V AC (switchable with [AC power supply voltage switch] on the bottom) Approx. 9VA when AC100V, Approx. 11VA when AC120V, Approx. 12VA when AC240V
 | 
| DC power supply | 10 to 30 V DC, 6 VA or less | 
| Operating temperature/humidity | -10°C to +50°C, within 20% to 85% RH, with no condensation | 
| Storage temperature/humidity | -20°C to +70°C, within 10% to 90% RH, with no condensation | 
| External dimensions | H143 x W49.5 x D252 mm (excluding projection) | 
| Weight | Approx. 1.4 kg |