Type of Battery  | Lithium-ion battery
  *United Nations Recommendations on the Transport of Dangerous Goods UN 38.3 safety tests: Passed *The UN 38.3 safety test certification and a certificate of Safe Transport of Chemical Goods (Test Report) issued by the Shanghai Research Institute of Chemical Industry, China: Received  | 
Input (DC IN)  | 10 to 30 VDC
  Connector: ECP.1S.302.CLL (LEMO) Compatible plug: FFA.1S.302.CLA** (LEMO) ** stands for the collet type, size number, etc.  | 
Output  | [Voltage]  10 VDC
  [Current]  Maximum: 2.5 A (Ambient temperature: 0 to 40°C)  Maximum: 1.5 A (Ambient temperature: -10 to 0°C)  Maximum: 2.0 A (Ambient temperature: 40 to 50°C)  | 
Operating Switch  | Battery-check switch
  [Press once]  Remaining battery power check  The battery-check LED indicates the remaining battery power.
  [Hold down (For approx. one second)]  Battery integrity check  The battery-check LED indicates the integrity.  | 
Indicator  | [Status LEDs: 1] During external-power operation: Lit up in blue During battery operation: Lit up in purple (The LED flashes purple when the remaining battery power is less than 30%.) When an error occurs: Flashes red
  [Battery-check LEDs: 3] ----When checking the remaining battery power (Lit up in blue) :  ●/●/● 100 to 70%  ●/●/- 69 to 30%  ●/-/- 29 to 0% * Flashes red at 5% or less
  ----When checking the integrity (Lit up in purple) :  ●/●/● Good  ●/●/- Caution  ●/-/- Replacement required
  ----When an error occurs:  Flash red (All 3)
  (● : Lit up, -: Off)  | 
Charging Time*1  | 
  | 
 When Charging the Battery Alone  | 4.5 h or less (Ambient temperature 10 to 40°C) 7.0 h or less (Ambient temperature 0 to 10°C)  | 
 When Supplying Power to the System and Charging the Battery at the Same Time  | 7.0 h or less (Ambient temperature 10 to 40°C) 10.0 h or less (Ambient temperature 0 to 10°C) *2  | 
Discharge Time  | During 2.5 A (25 W) output: 60 min or more During 0.5 A (5 W) output: 300 min or more * When using a new battery at an ambient temperature of 25 ±10°C
  (Reference) The power consumption of each unit is as follows:  CTRS-100A 3.5 W  CTRS-CDV010A 2.1 W  CTRS-SYNC010A 0.3 W Because the total power consumption in the case of analog measurement on 32 channels + the use of synchronization unit is approx. 18.5 W, operation for 80 minutes or more is possible.  | 
Operating Temperature  | 
  | 
 Charging  | 0 to 40°C  | 
 Discharging  | -10 to 50°C  | 
Operating Humidity  | 20 to 90% (Non-condensing)  | 
Storage Temperature  | -20 to 50°C * In the case of long-term storage for one month or more, avoid high temperatures or humidity, and store the unit at 40°C or less.  | 
Vibration Resistance  | 49.0 m/s2 (5 G), 5 to 200 Hz  | 
Shock Resistance  | 490 m/s2 (50 G), 11 ms or less, half sine wave  | 
Degree of Protection  | IP50 (IEC 60529)  | 
Dimensions  | 53.2 W × 92 H × 94 D mm (Excluding protrusions or protectors)  | 
Weight  | Approx. 570 g  | 
Compliance  | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)  | 
Utility Nuts  | Size: M4, 12 places  | 
Battery Pack Replacement  | Handled by Kyowa (Replacement by the user is not possible.)  | 
  | *1 If the battery is hot, it may stop charging before it is fully charged for safety reasons. This is not a malfunction. *2 Depending on the environment, charging may stop before the battery is fully charged for safety reasons. This is not a malfunction.  |