KYOWA
CCA-90A IEPE Accelerometer Card
Multi Signal Conditioner
Specifications:
Channels | 1 |
Measuring Targets | IEPE accelerometer Max. ±5000 mV (built-in amplifier type) *Charge converter is necessary when using a charge output accelerometer.
Recommended item: Fuji ceramics corporation, CAC1R0 |
Input | Format: Unbalanced input Sensor supply: Built-in constant-current supply (constant current: 4mA) Excitation voltage: Approx.24 V |
Frequency Response | 0.2 Hz to 50 kHz (Deviation: +1, -3 dB) |
Sensitivity Adjustment | Sensitivity is set in combination with a 4-digit INPUT switch and a 3-digit OUTPUT switch.
INPUT switch range: 20 to 5000 mV in 1-mV step OUTPUT switch range: 1.00 to 5.00 V in 0.01-V step Accuracy: Within ±(0.5% +5 mV) Range: ×1 to ×250 |
Fine Sensitivity Adjustment | ×0.4 to ×1 |
Calibration (DC CAL) | ±(1 to 5000 mV) Accuracy: Within ±0.3% FS |
LPF | Transfer characteristic : 2nd order Butterworth Cutoff frequencies : 300, 1k, 3k, 10k Hz and FLAT (5 steps) Amplitude ratio at cutoff point : -3 ±1 dB Attenuation : (-12 ±1) dB/oct. |
Distortion Rate | Within 1% (When ±5 V output) |
SN Ratio | 48 dBP-P or more (When 20 mV input and 5 V output) |
Stability | Temperature Zero point: Within ±0.5 mV per °C Sensitivity: Within ±0.05%/°C
Time Zero point: Within ±5 mV per 8h Sensitivity: Within ±0.5%/8h |
TEDS | Reads the sensor's TEDS information then displays it on the monitor meter to the value that converted the output voltage into the physical quantity. |
Output | Dual output, both BNC connector and integrated connector output the same signal.
Voltage output: ±5 V or more, when load resistance 5 kΩ or more Zero adjustment: -0.1 to 0.1 V or wider |
Output Off Function | Available |
Output Impedance | 2 Ω or less |
Withstand Voltage | 500 VAC for 1 min, between the following two positions respectively: Input and output Input and case Output and case |
Over Input Indication | LED lights up. |
Connector Type | Input connector: BNC Output connector: BNC |
Dimensions | 24 W ×96 H ×170 D mm (Excluding protrusions) |
Weight | Approx. 140 g |
Compliance | Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
Note | The specifications does not include the accuracy of the optional charge converter. |
|